MAG 8000 & MAG 8000 CT
24
SFIDK.PS.026.D9.02
FT/PDM Meter
Display Parameter/data type
Factory settings
Data range
number
version view
Fix parameter or meter data that
not
are
changeable
312
All
-
Flow direction totalizer 2
Reverse
forward, reverse or bi-directional net flow
313
All
-
Totalizer 2 changes date
PS3 production date and time
320
All
-
Verification mode enable
No
Yes / No
321
All
-
Calibration date
Calibration date
year-month-day T
hours:minutes:seconds
323
All
-
Calibration factor
Sensor related
324
All
-
Gain correction
Sensor related
325
All
-
Sensor offset
Sensor related
327
All
-
Adjustment Factor
1
-2 to 2
328
All
-
Low flow cut off
0.05%
0 to 9.9%
329
All
-
Filter time constant
5 Tau
1 to 1000
331
All
-
Excitation frequency limit
6.25Hz for advanced version
1/15Hz, 1/5Hz, 1.5625Hz, 3.125Hz,
and 1/15Hz for basic version
6.25Hz, 1/30Hz, 1/60Hz
332
All
-
Excitation frequency sensor limit Sensor related
6.25 Hz (DN25...DN200 (1"...8"))
3.125 Hz (DN250..DN600 (10"..24"))
333
All
-
Empty pipe detection enable
Yes
Yes / No
334
All
-
Empty pipe limit
25000 ohm = 20uS/cm
0 to 2.15*10^9
400
All
-
Output A enable
Yes
Yes / No
401
All
Se3
Pulse A function
Forward
Forward, Reverse, Forward net,
Reverse net
402
All
Se3
Amount per pulse A
Sensor related
0-1*10^10
403
All
-
Pulse width for pulse A
50 ms
5 ms, 10 ms, 50 ms, 100 ms, 500 ms
404
All
-
Output B enable
Yes
Yes / No
405
All
Se4
Pulse B function
Alarm
pulse, alarm, call-up
406
All
-
Pulse B direction
Reverse
Forward, Reverse, Forward net,
Reverse net
407
All
Se4
Amount per pulse B
Sensor related
0-1*10^10
408
All
-
Pulse width for pulse B
Sensor related
5 ms, 10 ms, 50 ms, 100 ms, 500 ms
420
All
M5
Device Communication Address
1
1 to 32
421
All
M6
Baudrate
19200
1200, 2400, 4800, 9600, 19200, 38400
422
All
M7
Parity
Even 1 stop
Even 1 stop, Odd 1 stop, None 2 stop,
None 1 stop
423
All
-
Interframe space
35
35 to 255
424
All
-
Response delay
5
1 to 50 ms
425
All
-
Reset communication driver
No
Yes / No
500
All
-
Latest service date
PS3 production date and time
year-month-day T
hours:minutes:seconds
501
All
-
Operating hours since power up
0
hours
502
All
-
Battery operating time
0
hours
505
All
-
Power supply
P
ower supply level
Battery or mains power
506
All
-
Numbers of power up
0
507
All
-
Battery power
MLFB depended
1 to 4
batteries
508
All
-
Battery change enable
No
Yes / No
509
All
Se1
Battery installation date
PS3 production date and time
year-month-day T
hours:minutes:seconds
510
All
Se2
Actual battery capacity
100%
100 to 0%
512
All
-
Excitations no.
0
513
All
-
Power status
0
0: Normal operation, 1: Battery alarm.
Actual battery capacity is below battery
alarm level (% of max capacity), 2: Too
low power (enters stand by mode), 3: As
value 1 and 2 together, 4: External power
gone, 5: As value 1 and 4 together, 6: As
value 2 and 4 together, 7: As value 1 and
2 and 4 together
514
All
-
Transmitter temperature
Actual degree celsius
540
All
-
Electrode impedance A
Measured values
0-185000 ohm
541
All
-
Electrode impedance B
Measured values
0-185000 ohm
542
All
-
Low medium impedance alarm
0
0 to 2.15*10^9
550
All
-
Coil current disable
No
Yes / No
551
All
-
Fixed flow mode enable
No
Yes / No
552
All
-
Fixed flow value
0
-1*10*10^9 to 1*10*10^9
553
All
-
Flow alarm limit
1000000000
0 to 1*10*9
560
All
-
Repair checksum
No
Yes / No
4. Commissioning
Summary of Contents for SITRANS F M MAG 8000
Page 60: ...MAG 8000 MAG 8000 CT 60 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 61: ...MAG 8000 MAG 8000 CT 61 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 62: ...MAG 8000 MAG 8000 CT 62 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 63: ...MAG 8000 MAG 8000 CT 63 SFIDK PS 026 D9 02 10 Approvals and certificates ...