MAG 8000 & MAG 8000 CT
31
SFIDK.PS.026.D9.02
Index 4
Active alarm
Fault information. Each number indicates a dedicated fault:
1
Insulation fault
2
Coil current fault*)
3
Preamplifier overload fault*)
4
Data base checksum fault
5
Low power warning
(alarm limits are configurable)
6
Flow overload > Q
max.
(Q4) (125% Qn) flow overload
7
Pulse output 1 overflow > PF [Hz] pulse output 1 overflow
8
Pulse output 2 overflow > PF [Hz] pulse output 2 overflow
9
Consumption interval warning
(alarm limits are configurable)
L
Leakage warning
(alarm limits are configurable)
E
Empty pipe / low conductivity - when enabled*
C
High conductivity/low impedance warning
(alarm limits are configurable)
d
High flow rate warning
(alarm limits are configurable)
*) Meter disables measurement to reduce power consumption during fatal faults.
Faults are indicated with the lowest number
first. The display on the left indicates 3 alarm
conditions: low power warning (5), leakage
warning (L), and empty pipe warning (E).
Faults 1 to 4 affect the performance of the meter
and remain active until the alarm condition
disappears. Faults 5 to d are warnings that will
disappear when the alarm condition has been
corrected and reset via communication inter-
face.
Fault evaluation and service guidelines are
made in the service section.
After all faults have disappeared the display
show total hours of faults until the meter is reset.
5. Operation
5.2 Operator menu
Index 1
Totalizer 1
Index 2
Totalizer 2
Index 3
Flow rate
Flow volume totalizer 1 (factory configured for
forward flow calculation).
Flow volume totalizer 2 (factory configured for
reverse flow). A negative value indicates re-
verse flow calculation.
The value of totalizer 2 can be reset to zero or
set to any value desired (example - replacing
an existing old meter).
Index 3 shows the actual flow rate. If a negative
value is indicated, flow is in a reverse direction.
MAG 8000 CT
The value of totalizer 1 or 2 can only be reset
to zero if the verification sealing is broken and
a hardware lock is mounted to the device.
MAG 8000
The value of totalizer 1 can be reset to zero or
set to any value desired (example - replacing
an existing old meter).
Summary of Contents for SITRANS F M MAG 8000
Page 60: ...MAG 8000 MAG 8000 CT 60 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 61: ...MAG 8000 MAG 8000 CT 61 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 62: ...MAG 8000 MAG 8000 CT 62 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 63: ...MAG 8000 MAG 8000 CT 63 SFIDK PS 026 D9 02 10 Approvals and certificates ...