Appendix F: Fuzzy Logging
What is Fuzzy Logging?
Fuzzy Logging
is a real time Data Compression technique.
Patent no. US 6,405,155 B2
,
developed
as an alternative to the more standard methods of recording data.
Paperless Recorders are primarily used for exception recording. They spend most of their
life trending and recording straight lines. Fuzzy Logging has been developed to improve the
efficiency of data storage, and is particularly effective in exception recording examples
where normal operation consists of generally static inputs.
Fuzzy Logging looks for straight lines in the data stream, in real time, whether they are hor-
izontal, climbing or descending. A straight line made up of say 10 points can be equally well
represented by 2 points, one at either end, the other 8 points are redundant. Fuzzy logging
works by creating straight lines in the data and discarding redundant points.
What’s it for?
……
To help the user in the trade off between
Scan Rate
,
Disk capacity
and
Recording
Time,
after all
the ‘Pie’ is only so big.
Fuzzy Logging has been developed to help
maximise all three sections, in effect increas-
ing the size of the ‘Pie’.
The result is a technique that delivers a host
of real world benefits over the more traditional
recording methods.
1. Disks take longer to fill – changed less frequently, less site visits.
2. Faster scan rates can be used for any given disk size – giving greater resolution on the
process.
3. Recording time can be extended.
4. Less hard disk memory required for archiving on the PC.
5. Quicker graphing of data.
6. Smaller data files for remote collection.
How does it work?
Fuzzy Logging does not log data points that form part of a straight line.
The technique is best illustrated by the diagram below:
The Recording 'Pie'.
Scan Rate
Storage Capacity
Recording Period
Summary of Contents for SIREC D200
Page 2: ......
Page 176: ...170 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 Firmware Credit System ...
Page 222: ...216 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Page 238: ...232 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Page 258: ...252 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Page 270: ...264 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Page 274: ...268 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 Modbus Function Codes ...
Page 284: ...278 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Page 285: ......