43-TV-25-35 GLO Iss.4 Dec 06 UK
217
A5E01001767-04
Section 11: Instrument Care and Maintenance
The
SIREC D200/D300/D400
range of recorders contains no parts serviceable by the
user, except for a fuse on the DC power supply, see
.
In the event of a recorder failure contact your nearest Service Department (or an author-
ised agent) to arrange for the return of the unit for repair.
Cleaning Instructions
The recorder’s touch screen can be temporarily disabled for cleaning, see
Cleaning the unit should be done with a soft lint cloth and warm soapy water or screen
cleaner recommended for use on a PC. Solvents and prolonged exposure to detergents
can cause damage to the front panel. It is recommended that any cloth used for cleaning
is damp but NOT wet, to avoid water collecting in the unit.
For Nema 4X /IP66 rated recorders (option) refer to the Nema 4X standards.
Backlights
1.
MTTF (Mean Time To Fail) is defined as the time at which 50% of a batch of backlights
remain in excess of half their original brightness i.e. a display has a greater than 50%
chance of being half its original brightness when used at 100% after:
• 55,000 hrs for the
SIREC D300
.
Maximum luminosity 400 cd/m2, (86,000hr if used at 80%).
• 43,000 hrs for the
SIREC D400
. Maximum luminosity 400 cd/m2, (67,000hr if used at 80%).
• 40,000 hrs for the
SIREC D200
.
Maximum luminosity 450 cd/m2, (62,500hr if used at 80%).
For this recorder the backlight cannot be changed separately so the display has to
be replaced.
Reducing the brightness can significantly improve the life of the backlight.
2.
At extended temperatures these times are reduced.
3.
The backlight is a consumable item, and will NOT last forever.
4.
A maintenance schedule of exchanging backlights should be every 5 years to maintain
the optimal disk clarity.
Operating Temperature
Operating temperatures are described in
“Specification Tables” on page 223
. Prolonged
operation at temperatures over 50 °C will accelerate degradation of the display backlight.
If the unit has been moved from a cold environment into a warm one, ensure that the unit
has reached a minimum temperature of 12 °C or is left to stand for 1 hour at room temper-
ature before applying power, to ensure no condensation remains in the unit.
Summary of Contents for SIREC D200
Page 2: ......
Page 176: ...170 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 Firmware Credit System ...
Page 222: ...216 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Page 238: ...232 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Page 258: ...252 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Page 270: ...264 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Page 274: ...268 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 Modbus Function Codes ...
Page 284: ...278 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Page 285: ......