3
Servo Ventilator 300/300A
Disassembling and assembling
E382 E380E 061 01 03 01
Siemens-Elema AB
87
Front panel film
Separate the panel section from the
control section as previously described.
Remove all front cover control knobs as
previously described. Remove only the
control knobs, not the complete controls.
Remove all panel holders (1) that secures
the front panel plate to the frame.
Lift off PC 1614 including the front panel
base plate (2).
Lift off the front panel film (3) from the
frame.
Panel interface board PC 1614
Separate the panel section from the
control section as previously described.
Disconnect all front panel control
connectors (1) from PC 1614.
Remove all nuts (2) holding PC 1614.
On newer units: Remove all panel
holders (3).
Carefully lift off PC 1614 (4) from the front
panel base plate.
When the panel interface board PC 1614
is separated from the front panel base
plate (e. g. as a spare or exchange part),
there are a number of components on
PC 1614 that can be easily damaged.
Therefore,
always
handle PC 1614 very
carefully.
3
2
1
1
300-A29X
1
2
4
300-L26X
2
2