126
Siemens-Elema AB
E382 E380E 061 01 03 01
Diagrams
Servo Ventilator 300/300A
8
Interconnection cable
The
INTERCONNECTION
CABLE
is the 2.95 m long
89 leads cable that connects the Control unit
with the Patient unit. Signal names and pin
numbers are listed below:
Connected at
Connected at
patient unit
control unit
N16
N50
Pin No.
Signal name
Pin No.
1
Dis Saf valve and 24 V L
1
2
O
2
cell connected L
2
3
–
3
4
SF 6 conc
4
5
CO
2
values
5
6
ET CO
2
conc
6
7
Insp flow
7
8
CO
2
conc
8
9
CO
2
trans connected
9
10
Insp time H
10
11
Receive address L
11
12
Technical error CO
2
exp flow
12
13
CPAP neonate mode H
13
14
Serial data to slave H
14
15
Exp detection H
15
16
Serial data to master H
16
17
Enable NEB valve H
17
18
Enable O
2
valve H
18
19
Enable AIR valve H
19
20
Zeroing insp valves H
20
21
Power up reset L
21
22
Exp time H
22
23
Mon dis insp valves L
23
24
Ref dis insp valves L
24
25
Reserve 2
25
26
Exp flow
26
27
Exp valve position
27
28
Exp tidal volume
28
29
Ext power overload
29
30
Pow sup out 1
30
31
Pow sup out 2
31
32
Pow sup out 3
32
33
Battery charging LED
33
34
–
34
35
Internal battery mode H
35
36
Valve code clock H
36
37
+24 V low power reg
37
38
Load valve code L
38
39
Ventilator on H
39
40
Serial valve code 1 H
40
41
Power on LED
41
42
Range sel D0
42
43
Adult range L
43
44
Range sel D1
44
Connected at
Connected at
patient unit
control unit
N15
N49
Pin No.
Signal name
Pin No.
1
+V
1
2
+V
2
3
GND
3
4
GND
4
5
GND
5
6
Sense GND
6
7
-V
7
8
-V
8
9
+5 VL
9
10
+5 V
10
11
+5 V
11
12
+5 V
12
13
+24 V
13
14
Sense +5 V
14
15
0 V (24)
15
16
0 VA
16
17
0 VA
17
18
0 VA
18
19
GNDL
19
20
Reserve 3
20
21
SGND
21
22
SGND
22
23
SGND
23
24
SGND
24
25
PS 4
25
26
–
26
27
Humidifier off L
27
28
Internal battery voltage
28
29
Power supply temp
29
30
External battery voltage
30
31
Inner shield (soldered)
–
32
–
32
33
Insp press
33
34
NEB flow ref
34
35
O
2
flow ref
35
36
AIR flow ref
36
37
NEB flow
37
38
O
2
flow
38
39
AIR flow
39
40
Valve position NEB
40
41
Exp press
41
42
Gas supply NEB press
42
43
O
2
conc
43
44
Valve position O
2
44
45
PEEP level set buff
45
46
Gas supply O
2
press
46
47
Barometer press
47
48
Valve position AIR
48
49
Airway flow
49
50
Gas supply AIR press
50
Summary of Contents for Servo Ventilator 300
Page 4: ...4 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 6: ...6 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 18: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 22: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 130: ......