Servo Ventilator 300/300A
Description of functions
E382 E380E 061 01 03 01
Siemens-Elema AB
73
15.3 O
2
concentration amplifier
This block, the
O
2
CONCENTRATION
AMPLIFIER
,
amplifies the signal from 11.3
O
2
CELL
.
The output signal is amplified to 90 mV/%O
2
at 1013 mbar (760 mmHg) ambient
barometric pressure. This output signal is
used in:
• 4
MONITORING
to monitor the O
2
concen-
tration where it is distributed to 1.5
O
2
CONCENTRATION
to display the measured
value on the front panel.
• 18.1
LINEARIZING
CPU
for compensation of
the
EXP
.
FLOW
value.
15.4 Valve zeroing
Zero offset is handled through:
• Timing control for the zeroing of the actual
gas flow and position signals within the
main blocks 8
INSPIRATORY
VALVE
UNIT
–
AIR
and 9
INSPIRATORY
VALVE
UNIT
–
O
2
when the
real actual gas flow is zero (normally
during pause time).
• Timing control for the zeroing of the actual
position signal within the main block 16
EXPIRATORY
VALVE
.
The timing control mentioned above is
executed in cooperation with the block 15.5
VALVE
ENABLING
.
15.5 Valve enabling
Valve enabling/disabling is handled through:
• Timing control (enabling) of the main
blocks 8
INSPIRATORY
VALVE
UNIT
–
AIR
and
9
INSPIRATORY
VALVE
UNIT
–
O
2
.
• Timing control (disabling) of the main block
16
EXPIRATORY
VALVE
.
15 Valve control
This main block consists of the board
PC 1622
VALVE
CONTROL
.
15.1 PEEP PID control
Controller block comparing actual
EXP
.
PRESSURE
signal from the main block 13
EXPIRATORY
PRESSURE
with desired
PEEP
LEVEL
signal according to front panel setting.
In order to somewhat decrease the
expiratory resistance at high expiratory
flows, the
EXP
.
FLOW
signal from the block
18.5
GAS
COMPOSITION
COMPENSATION
is used in
addition to the actual
EXP
.
PRESSURE
value in
such a way that a high
EXP
.
FLOW
value results
in a somewhat bigger opening of the
EXPIRATORY
VALVE
compared to when the
EXP
.
FLOW
value is small.
The actuating output signal is used as a
desired
POSITION
REF
signal for the block 15.2
POSITION
PD
CONTROL
.
15.2 Position PD control
Controller block comparing actual
EXP
.
VALVE
POSITION
signal from the block 16.3
EXPIRATORY
SOLENOID
WITH
POSITION
SENSOR
with desired
POSITION
REF
signal from the block 15.1
PEEP
PID
CONTROL
. The actuating output signal is
used as desired
EXP
.
VALVE
CURRENT
REF
for the
block 16.1
CURRENT
CONTROL
.
Summary of Contents for Servo Ventilator 300
Page 4: ...4 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 6: ...6 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 18: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 22: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 130: ......