Siemens-Elema AB
E382 E380E 061 01 03 01
Service procedures
Servo Ventilator 300/300A
Selecting the language and
barometric pressure unit
The following selections are made with the
switch SW1 on PC 1608
MONITORING
:
– The language of the text shown in the
”Alarms and messages” display.
– The unit of the actual barometric pressure
shown in the ”Alarms and messages”
display.
PC1608 must be removed from the control
unit to make SW1 accessible. See also
chapter ”
3. Disassembling and assembling
for detailed disassembling and assembling
instructions.
• Disconnect and prepare the ventilator as
described in chapter ”3. Disassembling
and assembling”, section ”
• Remove PC1608 from the control unit.
Selecting the Alarms and
messages language
This illustration shows how the switches 1,
2 and 3 should be set to select the different
available languages.
Switch settings not showed in this
illustration will normally give english
language.
Selecting the barometric pressure
unit
This illustration shows how the switch 4
should be set to select mmHg or mbar.
The actual barometric pressure is showed in
the ”Alarms and messages” display when
the touch pads ”Airway pressure” and
”Technical” are touched simulaneously
three times.
1
2
3
4
OPEN
1
2
3
4
OPEN
mbar
mmHg
300-A50X
1
2
3
4
OPEN
1
2
3
4
OPEN
OPEN
OPEN
OPEN
OPEN
1
2
3
4
1
2
3
4
1
2
3
4
1
2
3
4
English
Spanish
German
Italian
French
Swedish
300-A49E
Summary of Contents for Servo Ventilator 300
Page 4: ...4 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 6: ...6 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 18: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 22: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 130: ......