Siemens-Elema AB
E382 E380E 061 01 03 01
Servo Ventilator 300/300A
General
Unless stated otherwise, the information
in this Service Manual is valid for:
Servo Ventilator 300 (SV 300)
Servo Ventilator 300A (SV 300A).
The documentation for the Servo
Ventilator 300/300A consists of:
Operating Manual
Service Manual
Circuit Diagram
Spare Parts Catalogue
Reference Manual, Computer Interface
The Operating Manual is an indispens-
able complement to the Service Manual
for proper servicing.
In addition to the Important information
given here and in the related documents,
always pay attention to applicable local
and national regulations.
Responsibility for the safe functioning of
the equipment reverts to the owner or
user in all cases in which service or
repair has been done by a non-
professional or by persons who are not
employed by or authorized by Siemens,
and when the equipment is used for
other than its intended purpose.
Data on internal pressures in the Servo
Ventilator 300/300A are given in Pa (bar).
Airway pressures are given in cm H
2
O.
1 hPa = 1 mbar
1 mbar = 1 hPa
1 kPa = 10 mbar
1 mbar = 0.1 kPa
1 kPa = 0.01 bar
1 bar = 100 kPa
1 kPa
i
10 cm H
2
O
1 cm H
2
O
i
0.1 kPa
1 kPa
i
0.01 at
1 at ³ 100 kPa
1 kPa
i
0.01 kgf/cm
2
1 kgf/cm
2
i
100 kPa
1 kPa
i
0.01 kp/cm
2
1 kp/cm
2
i
100 kPa
1 kPa
i
0.145 psi
1 psi
i
6.9 kPa
Calibration and functional check
After any service intervention in Servo
Ventilator 300/300A, perform a
Calibration and Functional check
according to instructions in the
Operating Manual.
Hazard notices
Before disassembling or assembling the
Servo Ventilator 300/300A, make sure
that the:
Gas supply is disconnected.
Mains power cable is disconnected.
Mode selector is set to Ventilator off.
If the mode selector is set in any other
position, the internal battery will
supply power to the PC boards.
The back-up batteries are disconnec-
ted when the power section is open.
All gas conveying parts are cleaned
according to instructions in the
Operating Manual.
Service
When working with ESD sensitive
components, always use a grounded
wrist band and a grounded work surface.
Adequate service tools must always be
used.
Some parts in the Servo Ventilator 300/
300A are comprised by a spare parts
exchange system. Your local Siemens
representative supplies factory
calibrated and tested exchange parts as
well as other spare parts indicated in the
spare parts catalogue.
Important
Important
Summary of Contents for Servo Ventilator 300
Page 4: ...4 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 6: ...6 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 18: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 22: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 130: ......