Servo Ventilator 300/300A
Introduction
E382 E380E 061 01 03 01
Siemens-Elema AB
17
1
Inspiratory feedback control system (cont'd)
• A
CTUAL
VALUE
is the
INSP
.
PRESSURE
signal
from the
INSPIRATORY
PRESSURE
TRANSDUCER
• The actuating signal from the
PRESSURE
PID
CONTROL
block is used as a
FLOW
REFERENCE
signal going to the
MIXER
.
The rest of the controlling system, including
compensating elements, is the same as in
flow regulation.
Pressure regulation
Inspiratory pressure regulation is used only
during inspiration time in all modes except
when flow regulation is used as mentioned
previously.
When PRESSURE regulation is used, the
shadowed main forward path includes the
PRESSURE
PID
CONTROL
block in the
REFERENCE
&
TIMING
main block:
• D
ESIRED
VALUE
is the
PRESSURE
REFERENCE
signal from (the microprocessor module
MODE
CPU
within) the main block
REFERENCE
&
TIMING
Pressure
regulation
DESIRED PRESS.
VALUE
ACTUAL PRESS. VALUE
POSITION
SENSOR
INSPIRATORY
VALVE
CONTROLLED SYSTEM
VALVE
SOLENOID
PRESSURE
TRANSDUCERS
FLOW
ACTUAL CURRENT VALUE
FLOW CONTROL
INSP. VALVE POSITION
ACTUAL
SUPPLY
PRESS.
VALUE
GAS
FLOW
VALVE POSITION
CALCULATOR
REFERENCE
& TIMING
PRESS.
PID
CONTR.
FLOW
REF.
MIXER
DESIRED
POSITION
VALUE
CURRENT
CONTROL &
CURRENT
POWER
POSITION PD
CONTROL
FLOW PI
CONTROL
DESIRED
CURRENT
VALUE
ACTUAL POSITION VALUE
GAS
FLOW
REF.
ACTUAL
GAS FLOW
VALUE
DESIRED
GAS FLOW
VALUE
GAS
SUPPLY
PRESS.
INSP. PRESSURE
300-A10E
PRESSURE REF.
INSPIRATORY
CONTROL
Summary of Contents for Servo Ventilator 300
Page 4: ...4 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 6: ...6 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 18: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 22: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 130: ......