3
88
Siemens-Elema AB
E382 E380E 061 01 03 01
Disassembling and assembling
Servo Ventilator 300/300A
Front panel controls
• Mount the control (1) on the front panel
base plate and connect the connector (2)
to the PC board.
• On older units: Check that there is an O-
ring (3) mounted on the knob.
On newer units: The knob pointer ring is
made of a flexible rubber material that
replaces the function of the O-ring.
The O-ring is not used.
• Put the knob on the control shaft.
• Adjust the knob so that the knob pointer
(4) aligns with the indication mark (5). The
indication mark is a small dot or a thin
guide line.
• Tighten the nut (6) slightly.
• Check that the knob pointer still aligns
with the indication mark.
2
1
300-A30X
Assembling the panel
section
The procedure for assembling the panel
section is the reverse of the disassembly
procedure previously performed.
The following points must be noted during
assembling:
Panel interface board PC 1614
• Carefully mount PC 1614 on the front
panel base plate. Make sure that the PC
boards indicator LEDs and touch sensors
fit into the corresponding holes in the front
panel base plate.
0.2 mm
3
4
5
6
7
8
9
300-A31X
Summary of Contents for Servo Ventilator 300
Page 4: ...4 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 6: ...6 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 18: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 22: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 130: ......