Service procedures
Servo Ventilator 300/300A
Replacing the fuses
Fuses F1, F2, F3 and F4
Fuses F1 – F4 are mounted on PC 1618
POWER
SUPPLY
inside the power section.
• Disconnect and prepare the ventilator as
described in chapter ”3. Disassembling
and assembling”, section ”
• Disassemble the power section as
described in chapter ”3. Disassembling
and assembling”, section ”
• Replace blown fuses with new fuses as
listed in the table below:
Fuses
Purpose
F1: F 6.3 A
Limits the total +24 V supply current
F2: F 1.6 A
Controls the +24 V
FUSED
voltage
F3: F 6.3 A
Internal battery connector fuse
F4: F 6.3 A
External battery connector fuse
• Assemble the power section.
Fuses F11 and F12
• Disconnect the mains power cable from
the ventilator.
• Open the cover (1).
• Pull out the fuse holders (2).
• Replace blown fuses with new fuses as
listed in the table below:
Mains power
Selector
Fuses
voltage
setting
F11/F12
100 V (±10%)
100 V
T 3.15 AL
120 V (±10%)
120 V
T 3.15 AL
220 V (±10%)
220 V
T 1.6 AL
230 V (±10%)
220 V
T 1.6 AL
240 V (±10%)
240 V
T 1.6 AL
• Close the cover (1). The selected voltage
is shown in the cover window.
1
F11
F12
300-L28X
2
Summary of Contents for Servo Ventilator 300
Page 4: ...4 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 6: ...6 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Page 18: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 22: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Page 130: ......