7ML19985GG03
Pointek CLS500 – INSTRUCTION MANUAL
Page 67
m
mmmm
Appendix A
To change from menu 10 to menu 00:
1. Set the rotary switch to 0.
2. Press the RED (+) button for about a second. The LCD briefly displays : M 00
followed by : SEL 0, to indicate that menu 00 is selected. When the button is
released the LCD displays the current PV value. No left-arrow is displayed at the
top left corner of the LCD in menu 00.
To change from menu 00 to menu 10:
1. Set the rotary switch to 0.
2. Press the BLUE button for more than one second. The LCD briefly displays : M 10
followed by : SEL1, to indicate that menu level 1 is selected. Then the LCD
displays PV, and a left-arrow is visible in the top left corner of the LCD, indicating
menu level 1.
Display the Highest / Lowest Recorded Value
1. Set the rotary switch to 2. The Highest / Lowest recorded values for TV currently
selected are displayed alternately.
2. Press the RED (+) button to select the Highest recorded value for display,
or: Press the BLUE (–) button to select the Lowest recorded value for display,
3. Press both buttons simultaneously for more than one second to reset the
recorded values back to the dynamic value of this TV. (This will also occur after a
reset [power-down] of the device.)
Select the Transmitter Variable (TV) for the Primary Variable (PV).
Menu
Item
Affected
by:
Rotary
Switch
Position
Left
Arrow
Description
Values
00
01
0
Off
Dynamic Value (PV)
Units (or % of range
a
)
selected in menu 01
10
On
a.
Available but not useful for Pointek CLS500.
Menu
Item
Rotary
Switch
Position
Left Arrow
Description
02
2
Off
Highest / Lowest Recorded Value
Menu
Item
Rotary
Switch
Position
Left
Arrow
Description
Values
01
1
Off
Transmitter Variable
selection for PV
0 * TVO (units are pF)
1
TV1 (units are user definable
only via HART)
P
TV0 (values displayed as %)