Page EN-6
Pointek CLS200/300 (Digital) – QUICK START MANUAL
7ML19985XA82
mm
mm
m
E
n
glis
h
Power connection to screw terminals (standalone unit)
1.
Loosen the lid clip and unscrew the lid
of the enclosure.
2.
Unscrew and lift up the digital display.
3. Connect the wires to the terminals:
polarity is not important. (Terminal is
removable.)
4. To use the Alarm Output, connect the
wires of an optional input to the Alarm
Output terminals: polarity is not
important. (Terminal is removable.)
5. Ground the instrument according to
local regulations
1
.
6. Tighten the gland to form a good seal.
7.
Replace the digital display.
8. To adjust the transmitter locally,
use the keypad. (See
Setup using 3-
button keypad
on page 7 and the
Quick Reference table on page 8).
After adjustment, replace the
enclosure lid and secure the lid clip.
Operation
Digital display
Selected in Mode 13: displays either the measured value, logical level, mode number and
numerical value for the selected mode, or electronics temperature.
Measured value display
• default display at startup
• value (in counts
2
) when Sensor (2) selected
3
1.
The usual PROFIBUS PA recommendation is to ground the shield on both the device side and the cable
side. In some cases it may be preferable to ground one side only, to avoid ground loops.
2.
A dimensionless value generated from the inverse of frequency.
3.
Switch function (open/closed) depends on setting at Mode 24.
lid clip
digital display
removable
terminal block:
for PROFIBUS
PA/power
connection
removable
terminal block
for alarm output/
solid-state
switch
gland
power
cable
white
power cable
reed
contact
white
red (+)
black (–
)
orange
(f)
PROFIBUS PA
PA+
PA–
sensor
alarm output
solid-state switch
test
input
bar graph
sensor uncovered
down arrow
(switch open
3
)
Summary of Contents for Pointek CLS200
Page 1: ...Capacitance Switches Quick Start Manual 09 2010 Pointek CLS200 CLS300 Digital ...
Page 19: ...Page EN 18 Pointek CLS200 300 Digital QUICK START MANUAL 7ML19985XA82 mmmmm English Notes ...
Page 37: ...Side DA 18 Pointek CLS200 300 Digital KVIKSTART MANUAL 7ML19985XA82 mmmmm Dansk Notater ...
Page 55: ...Seite DE 18 Pointek CLS200 300 Digital KURZANLEITUNG 7ML19985XA82 mmmmm Deutsch Notizen ...