11
Water recharge:
OFF
Not used:
OFF
Not used:
OFF
SP. hum. CTRL
OFF
A7
(1)
(2)
(3)
(4)
1) When it is switched ON, the condensate water is recycled
When it is switched OFF, the condensate water is drained off.
2) Not used
3) Not used
4) When it is switched ON, the humidity slope can be controlled.
SHIFT
ENTER
If you enter the wrong number, you will still have to press ENTER and then re-digit
the number.
Aux 1:
OFF
Aux 2:
OFF
Aux 3:
OFF
Aux 4:
OFF
A8
(1)
(2)
(3)
(4)
When it is switched ON, the ABCD
contacts on the DIGITAL IN/OUT
connector are closed (see attached
electrical diagram and the descrip-
tion of the connector in the cham-
ber instruction handbook).
Aux 5:
OFF
Aux 6:
OFF
Aux 7:
OFF
Aux 8:
OFF
A9
(1)
(2)
(3)
(4)
SHIFT
ENTER
If you enter the wrong number, you will still have to press ENTER and then re-digit
the number.
SHIFT
ENTER
If you enter the wrong number, you will still have to press ENTER and then re-digit
the number.
TEST START-UP
K1
TO VISUALIZE
• To start the test
• The led light switches on the machine starts up and reaches the set tempera-
ture and humidity values: 20
°
C, 60% RH and remains in operation along these
values.
Both the set rising and descending values are reached at maximum speed.
K2
To switch off (led light switched off)
• Press
Data measured inside the chamber are
visualized.
The chamber reaches these values and
maintains them.
Temp. meas.
.......
°
C
Humid. meas.
.......%
In some versions of the programme the
writing in points (1), (2), (3), are
replaced by:
Dedicated contact 5
Dedicated contact 6
Dedicated contact 7