28
90
95
To set segment
7 (
°
C)
Temperature 30
°
C
Duration 2h - Maintenance
hh : mm : ss
Durat.
2 : 00 : 0
Wait duration
: ON
H1
Channel 1
Control
: ON
Wait set point
: OFF
Max speed
: OFF
Set point
: 30.00
°
C
Grad.
: 0.00
°
C/min
Near set
: 1,00
°
C
H2
H3
Because the same temperature value
is maintained
OP2605-110
Channel 2
Control
: ON
Wait set point
: OFF
Max speed
: OFF
H4
H5
OP2605-100
Set point
: 95.00 RH %
Grad.
: 0.00 RH %/min
Near set
: 3,00 RH %
to save the programmed segment
Save segment
: SAVE
H8
EDIT PROGRAM
< ESC > Prev. menu
Segment N. 8
Wait for this number to appear
7
Water recharge:
OFF
Not used:
OFF
Not used:
OFF
SP. hum. CTRL.
ON
To improve the system performance
(experimentally determined)
H6
To set segment
7 (% RH)
Humidity from 95%
Duration 2h - Maintenance
55
30
55
7
Unreal gradient which enables maxi-
mum speed to be obtained
H7
SHIFT
ENTER
This page is visualized
Save segment
: OFF
Total segment
: 0
Confirm progr
: OFF
When this page is visualized, if the previ-
ously described operations have been
carried out, the segment has been saved
and therefore must not be saved again.
Move on to the next step.
WARNING!
If you have accidentally saved the same segment twice, correct the number on this page.
• Press