29
95
The following four segments are the same as 4,5,6,7 and therefore a repetition loop will be set
later on.
When segment 7 has been programmed, the procedure will pass on to segments 8,9,10,11 and
then a repetition loop will be inserted.
To set segment
8 (
°
C)
Temperature from 30
°
C to 95
°
C
Duration 1.5h - slope -gradient 0.70
°
C/min
hh : mm : ss
Durat.
1 : 30 : 0
Wait duration
: ON
L1
Channel 1
Control
: ON
Wait set point
: OFF
Max speed
: OFF
Set point
:95.00
°
C
Grad.
: 0.70
°
C/min
Near set
: 1,00
°
C
L2
L3
Because the same temperature value
is maintained
OP2605-130
Channel 2
Control
: ON
Wait set point
: ON
Max speed
: OFF
L4
L5
OP2605-120
Set point
: 95.00 RH %
Grad.
: 0.00 RH %/min
Near set
: 3,00 RH %
to save the programmed segment
Save segment
: SAVE
L8
EDIT PROGRAM
< ESC > Prev. menu
Segment N. 9
Water recharge:
OFF
Not used:
OFF
Not used:
OFF
SP. hum. CTRL.
ON
To improve the system performance
(experimentally set)
L6
To set segment
8 (% RH)
Humidity 95%
Duratation 2,h - Maintenance
95
30
8
Unreal gradient which enables maxi-
mum speed to be obtained
L7
Segment 8, from 30
°
C to 110
°
C is divided into 3 segments (8,9,10 as humdity must be
switched off at 95
°
C)
8
Wait for this number to appear
SHIFT
ENTER
This page is visualized
Save segment
: OFF
Total segment
: 0
Confirm progr
: OFF
When this page is visualized, if the
previously described operations have
been carried out, the segment has
been saved and therefore must not be
saved again.
Move on to the next step.
• Press
repeatedly
WARNING!
If you have accidentally saved the same segment twice, correct the number on this page.
Summary of Contents for OP7
Page 3: ......