HT and single tank error diagnosis
6 - 5
Siemens-Elema AB
SPR8-215.898.01
Page 5 of 12
MOBILETT Plus / - E/ - M
Solna, Sweden
Rev. 05
01.02
SPS-UD
Exposure values: 40 kV 5 mAs, HT 8 kV/div (200 mV/div),
I
inverter-single tank
100 A/div (only the end of the exposure is shown)
.
Fig. 2
HT and inverter current, inverter and single tank normal
Inverter
current
HT on
D1