Mechanical systems
10 - 19
Siemens-Elema AB
SPR8-215.898.01
Page 19 of 28
MOBILETT Plus / - E/ - M
Solna, Sweden
Rev. 05
01.02
SPS-UD
•
Remove the all arm covers and the column covers.
•
Relieve the tension in the spring pack by loosening the adjustment screw above the tilt
bar.
•
Disconnect all cables to the collimator.
•
Remove the collimator.
•
Disconnect all electrical connections to the single tank.
•
Remove the supporting hub in the rotation axis of the fork.
•
Remove all cable clamps and cable ties on the arm and the chassis.
•
Disconnect the single tank and the fork from the arm.
•
Disassemble the fork so the cables can be removed.
•
Disconnect the chassis connections and remove the cables (Fig. 22).
Fig. 22
Danger of injury!
The arm may move uncontrollably when the weight of the fork
and single tank is removed.
Ensure that the arm is properly secured to the unit foot with the
rope.
Demount
Remount
WARNING