Capacitor bank
7 - 3
Siemens-Elema AB
SPR8-215.898.01
Page 3 of 6
MOBILETT Plus / - E/ - M
Solna, Sweden
Rev. 05
01.02
SPS-UD
•
Measure the voltage between the + and - bars; the voltage must be less than 10 V.
•
Note the cable connections to the capacitor bank (Fig. 3 for MOBILETT Plus/ Plus M and
Fig. 4 for MOBILETT Plus E)
Fig. 3
MOBILETT Plus/ Plus M
Fig. 4
MOBILETT Plus E
ll
MOB00494
- bar (neg)
+ bar (pos)
MOB00468
- bar (neg)
+ bar (pos)