Service programs
Pressing the lamp button stores the currently displayed values and also exits the program
P14.
P15 - Upper mAs limit
5
The program enables you to set the upper mAs limit. The available range is
0.5 - 200 mAs for MOBILETT Plus/Plus M and 0.5 - 100 for MOBILETT Plus E. The maxi-
mum mAs value depends on the actual kV but can be additionally limited. When selecting
an mAs limit value, all prior higher mAs values will be blocked and vice versa.
To change or check the highest available mAs, activate the program with the lamp button.
The display shows the selected mAs limit, change with mAs + or mAs - buttons.
Example:
Press the lamp button to store and currently displayed values and to exit P15.
P16 - Show the exposure counter
5
This program is activated by the lamp button and shows the actual number of exposures.
Example: 1
Example: 2
Exit the program by pressing the lamp button.
Always check that the values selected correspond to the desired
values after exiting the service program.
kV
mAs
SEt 100 = The mAs value is limited to 100 mAs
kV
mAs
599 = The number of exposures is 599.
kV
mAs
1
15055 = The number of exposures is 115.055
The exposure counter can not be cleared via any of the service
programs.The value is stored in the CMOS memory.
If the backup battery is depleted, the value could be lost.
NOTE
NOTICE