MOBILETT Plus / - E/ - M
SPR8-215.898.01
Page 6 of 10
Siemens-Elema AB
Rev. 05
01.02
SPS-UD
Solna, Sweden
5 - 6
Service programs
150-250-299-325-337-345-348-351-353 V. At the 353 V level, the charging will continue
for another 48 minutes to decrease and stabilize the capacitors internal leakage current.
If an error is detected during the formation program, the program is aborted. When the
program is aborted, there is an audible beep and a message is displayed:
If the formation ends successfully, the display will show:
Duration: Approx. 100 minutes.
P07 - Query filament values from the battery-backed RAM
5
The kV-values are displayed in the "kV" field. The corresponding filament values are dis-
played in the "mAs" field. Pressing the "kV +" button will display the next higher kV and
corresponding filament value and vice verse with "kV -".
The program starts at 40 kV in the table containing the filament values. The program
shows only the selected number of kV steps (25 or 49 for MOBILETT Plus/Plus M and 24
or 47 for MOBILETT Plus E).
The kV is changed in steps by briefly depressing the "kV + or kV -" buttons. By keeping
the buttons pressed, the kV value will be continuously stepped up or down.
The generator filament values are self-adjusted to produce a kV value with minimum devi-
ation. If the software was optimized, a filament value for a specific kV value was indicated
with the letter "A" in the extreme right position on the mAs display.
P08 - Display the message history
5
The history contains two parts. The first is the appearance number and the second the
actual message. The appearance number will be displayed in the "kV" field. The corre-
sponding message will be displayed in the "mAs" field. Pressing the "kV +" button will dis-
play the next higher number and corresponding message and vice verse with "kV -". The
program starts by showing the latest message in the history. If there are no messages the
display will show:
The latest message has the highest number, for example:
To exit P08, use the "lamp" button.
kV
mAs
P06 Err"xx" = See actual "xx" in separate message list
kV
mAs
P06 PASS = Capacitor formation completed
kV mAs
0
Error = Message history empty
kV mAs
20 CAL13 = There are total 20 new messages, number 20 is "CAL 13".