Service and maintenance
6.4 Conversion of MicroSAM basic device AS09
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
97
The gas chromatograph has now been configured for type A analysis modules.
Note
Parameter settings
The parameters on the CD are appropriate for basic devices up to AS08. The parameters
include the settings for the solenoid valves MV1 to MV4. MV1 is preset in the software to
"FlushLoop".
From AS09 onwards, the basic device contains an additional solenoid valve MV5, and MV1
and MV5 are preset to "FlushLoop_1" and "FlushLoop_2" respectively.
When reconfiguring to a gas chromatograph with type A analysis modules, you must
therefore rename "FlushLoop" to "FlushLoop_1".
You do not require "FlushLoop_2" for a gas chromatograph with a type A analysis modules.