
Service and maintenance
6.4 Conversion of MicroSAM basic device AS09
MicroSAM
94
Operating Instructions, 06/2012, C79000-G5376-C560-07
6.4
Conversion of MicroSAM basic device AS09
A solenoid valve MV5 is fitted in the MicroSAM basic device of release version 09 (AS09).
You can also prepare the MicroSAM basic device as a GC which can be operated with
analysis modules of type A.
6.4.1
Preparation for use of analysis modules of type A
Procedure
1.
Terminate measuring mode of the device.
2.
Deactivate the detectors using the operating software.
3.
Save the configuration in the flash memory.
4.
Switch off the device.
5.
Shut off the carrier gas and sample gas supplies.
6.
Open the two housing covers.
7.
Deinstall the analysis module type D.
8.
Install the analysis module type A.
9.
Remove the RSP module.
Summary of Contents for MicroSAM
Page 1: ...Process Gas Chromatography Process Gas Chromatograph MicroSAM Operating Instructions 03 2012 ...
Page 2: ......
Page 8: ...Table of contents MicroSAM 6 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 10: ...Purpose of this documentation MicroSAM 8 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 106: ...Examples MicroSAM 104 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 114: ...Dimension drawings MicroSAM 112 Operating Instructions 06 2012 C79000 G5376 C560 07 ...