
Service and maintenance
6.4 Conversion of MicroSAM basic device AS09
MicroSAM
96
Operating Instructions, 06/2012, C79000-G5376-C560-07
6.4.2
Restoring of data for AS09 type A
Requirement
Knowledge of the Maxum EZChrom und Maxum System Manager operating software.
Procedure
1.
Start up the device.
2.
Start the EZChrom program (start bar Start ➜ Programs ➜ Maxum System Tools ➜
Maxum EZChrom).
3.
Start the "Open" function (menu "File" ➜ Menu command "Method" ➜ "Open").
4.
Download the AXX.met file from the CD "Parameter Backup CD".
5.
Click in the toolbar of Maxum EZChrom on the "Instrument Setup" symbol, and select the
"Valves" menu.
6.
Click on "FlushLoop" in the "Name" column, and change the name in the drop-down list to
"FlushLoop_1".
7.
Repeat step 6 for all other entries with the name "FlushLoop".
8.
Change the remaining parameters according to your application.
9.
Save the configuration (menu "File" ➜ Menu command "Method" ➜ "Save As").
10.
Export the saved data to the GC (menu "File" ➜ Menu command "Method" ➜ "Export...").
Summary of Contents for MicroSAM
Page 1: ...Process Gas Chromatography Process Gas Chromatograph MicroSAM Operating Instructions 03 2012 ...
Page 2: ......
Page 8: ...Table of contents MicroSAM 6 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 10: ...Purpose of this documentation MicroSAM 8 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 106: ...Examples MicroSAM 104 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 114: ...Dimension drawings MicroSAM 112 Operating Instructions 06 2012 C79000 G5376 C560 07 ...