Technical data
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
107
Oven
Number/type
1/isothermal
Purging with N
2
possible
Dimensions
160 x 10 (diameter x height in mm)
Heat output
20 W
Tripping of thermal fuse
≥ 159 °C
Temperature range
60 … 155 °C
Temperature stability
± 0.1 K (60 to 155 °C)
Temperature accuracy
± 3 K (60 to 155 °C)
Retention time variations per 10 °C change in ambient
temperature
Approx. 0.3%
Heating-up period from 30 to 100 °C
10 minutes
Tripping of thermal fuse
ETC3
≥ 159 °C
ETC4
≥ 80 °C
Columns and gases
Column type
Capillary columns 0.15 to 0.25 mm ∅
internal diameter
Column switching
Multidimensional chromatography with backflushing and cut in
live system
Multifunctional membrane valve
For injection and backflushing
Gas connections
Swagelok 1/8"
Pressure controller
Max. 4 single-channel electronic pressure controllers
Solenoid valves for control of multi-function membrane
valve
2 NC contacts, 2/3 NO contacts
Carrier gas
H
2
, N
2
, He, Ar (application-dependent)
Gas purity (minimum requirement)
≥ 99.999% (5.0)
Particulates
< 1 μm
Required filtration
Degree of separation 99.99 % for 1 μm particles
Consumption
10 to 60 ml/min (application-dependent)
Input pressure
500 …700 kPa (absolute)
Instrument air
Not required
Summary of Contents for MicroSAM
Page 1: ...Process Gas Chromatography Process Gas Chromatograph MicroSAM Operating Instructions 03 2012 ...
Page 2: ......
Page 8: ...Table of contents MicroSAM 6 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 10: ...Purpose of this documentation MicroSAM 8 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 106: ...Examples MicroSAM 104 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 114: ...Dimension drawings MicroSAM 112 Operating Instructions 06 2012 C79000 G5376 C560 07 ...