Service and maintenance
6.4 Conversion of MicroSAM basic device AS09
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
95
10.
Interchange the plugs to the solenoid valves MV3 and MV5.
Procedure: Preparation for use of analysis modules of type A
View of EPC module
①
Disconnect the solenoid valve MV5 (plug number 5).
①
Disconnect the solenoid valve MV3 (plug number 3).
③
Connect the cable number "3" to the solenoid valve MV5. The cable number "5" is not
required in the configuration of type A.
11.
Reinsert the RSP module.
12.
Close the housing.
The basic device AS09 is now prepared for use of analysis modules of type A.