Troubleshooting
10.3 Test Facilities Graph Screen
FUG1010 IP65 NEMA 4X
Operating Instructions, 01/2013, A5E02951519-AC
155
Signal Masking Function: (Hot Key 7)
Under conditions of extremely low signal amplitude, a noise spike associated with the flow
meter receive signal window may be present on the extreme left side of the graph display. If
this spike is large enough it may interfere with the signal detection routines.
1.
To eliminate this noise from the signal processing routines, press the <7> key to invoke
the [Mask Set #] command, then press the <+> key until the noise is no longer present in
the receive waveform.
2.
Press <0> to exit this command.
Hold Set Function: (Hot Key 8)
The [Hold Set #] command is used to set the Hold Set number higher if intermittent mis-
registration occurs. Press the <8> key on the keypad to invoke this function.
Table 10- 2 Description of Graph Screen Text Display Parameters
Screen Text Parameters
Menu List Item
Description
F
Flow
Measured flow rate in selected flow
units.
VS
Vs m/s
Sound Velocity in meters per second.
[ ]
Display Metrics
Represents the digital sample position
of the receive window.
Correlated plot
Displays the receive waveform in its
proper superposition or registration.
The true delta time will be displayed by
NOT selecting "Correlated Plot".
Centroid Mark
Indicates with a large vertical marker
the peak energy of the receive
waveform.
D
Damping
Displays the minimum and maximum
digital damping exponent along with the
active damping exponent.
Tn
Tn (usec)
Receive signal transit time in
microseconds.
dT
DeltaT (nsecs)
Transit time delta (difference) in
nanoseconds.
S
Signal Strength
Displays %Valc (amplitude), %Vaer
(aeration factor) and numeric ALC.
Signal-to-Noise Ratio
Indicates the signal to noise ratio of the
receive signal. Increased damping will
increase the S/N ratio as the
asynchronous noise reduces.
SN
Envelope
Percentage change of the signal from
Initial Makeup conditions.
Summary of Contents for FUG1010 IP65 NEMA 4X
Page 9: ...Table of contents FUG1010 IP65 NEMA 4X 8 Operating Instructions 01 2013 A5E02951519 AC ...
Page 196: ......
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ...21614 C 1011NFPS 7 ...
Page 238: ...1011NFPS 7 C 21614 ...
Page 239: ...1011NFPS 7 C 21614 ...
Page 240: ...21614 C 1011NFPS 7 ...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ...21614 C ...
Page 247: ......
Page 248: ......
Page 249: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 250: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 251: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 252: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 266: ...Glossary FUG1010 IP65 NEMA 4X 200 Operating Instructions 01 2013 A5E02951519 AC ...
Page 270: ...Index FUG1010 IP65 NEMA 4X 204 Operating Instructions 01 2013 A5E02951519 AC ...