Functions
7.4 Zero Flow Adjust Menu
FUG1010 IP65 NEMA 4X
Operating Instructions, 01/2013, A5E02951519-AC
109
Using ReversaMatic
If site conditions do not permit stopping the flow rate at the mounting location, and you do
not know the current flow rate, then you can use the ReversaMatic routine to establish the
zero flow level. You should perform the ReversaMatic procedure as quickly as possible to
ensure that the flow rate remains constant throughout the procedure.
To invoke ReversaMatic:
1.
To access the [Zero Flow Adjust] option list press <Right Arrow>.
2.
Move the cursor to [ReversaMatic]. Press <ENTER> to invoke the routine.
3.
The flow meter begins to measure the positive flow rate. "Positive" flow refers to flow
moving from upstream sensor location to the downstream sensor location. Note top
prompt line shows: Reversamatic Action
4.
Upon completion, the flow meter beeps and the display screen shows: Reverse Sensors /
Press <ENTER>
5.
Now remove then remount the upstream and downstream sensors in their reversed
positions. Mount the Up sensor (without removing its cable) in the Down sensor/cable
location. Mount Down sensor with its cable in the Up sensor/cable location. When
remounting the sensors, couple them to the pipe properly. Press <ENTER> (after re-
installing the sensor).
6.
The flow meter measures the negative flow rate briefly, then beeps and repeats the
prompt: Reverse Sensors / Press <ENTER>
7.
Now remount the sensors for normal operation (in their original orientation). When
remounting sensors, couple them to the pipe properly. Press <ENTER> (after re-installing
the sensors).
This completes the ReversaMatic procedure. The system’s zero accuracy will be very close
to that obtainable using the Actual Zero method, providing flow remained constant during this
procedure.
NOTICE
Preventing Flow Mis-Registration
A caution on the use of upper and lower flow limits (used to prevent flow mis-registration)
prior to using the Reversal Zero technique (ReversaMatic): If the negative flow rate that the
flow meter reads in the step during which the sensors are reversed is more negative than
the lower flow limit, the meter will re-register positive and the Reversal Zero cycle will thus
be corrupted.
Therefore, postpone the installation of upper and lower flow limits until the reversal zero
procedure is executed successfully. For pipes that combine large diameters with very high
flow velocities, it may be necessary to move the upper and lower flow limits out of the way
until the reversal zero is completed. Moreover, pipes of this size frequently have excellent
intrinsic zero performance and may not even need zeroing.
Summary of Contents for FUG1010 IP65 NEMA 4X
Page 9: ...Table of contents FUG1010 IP65 NEMA 4X 8 Operating Instructions 01 2013 A5E02951519 AC ...
Page 196: ......
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ...21614 C 1011NFPS 7 ...
Page 238: ...1011NFPS 7 C 21614 ...
Page 239: ...1011NFPS 7 C 21614 ...
Page 240: ...21614 C 1011NFPS 7 ...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ...21614 C ...
Page 247: ......
Page 248: ......
Page 249: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 250: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 251: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 252: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 266: ...Glossary FUG1010 IP65 NEMA 4X 200 Operating Instructions 01 2013 A5E02951519 AC ...
Page 270: ...Index FUG1010 IP65 NEMA 4X 204 Operating Instructions 01 2013 A5E02951519 AC ...