
Operation
5.4 Operating functions
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
73
5.4.5.4
On/off functions (function 52)
52 On/off config. O
2
Auto range (Fct.4O)
Store function
(Fct.77)
...continue
Figure 5-36 On/off functions
With this function, you can easily switch other functions on/off, such as the ones listed in the
above function screen.
Thanks to this simplified operation, these functions do not have the longer paths through
various menus. For better orientation, the function numbers have also been specified.
Activated functions are marked by "■", deactivated ones by "□". Using the fifth softkey
("...continue"), jump to the next function screen with further functions.
Function 52 is used to switch the following functions on and off:
Table 5- 3
Functions switched on/off by function 52
Description
Fct. No.
Remarks
Automatic measuring range switching
40
Measured-value memory
77
Temperature recompensation of zero point
86
Temperature recompensation of span
86
Signal tolerance violation
78
Limit monitoring 1
51
Limit monitoring 2
51
Limit monitoring 3
51
Limit monitoring 4
51
Suppress negative measured values
70
Automatic calibration
24
Only with supplementary
electronics
Total calibration
23
Block logbook
60
Fault / WA / CTRL MAMUR
72
Suppress negative measured values for display
55
Besides the functions listed in the table "Functions switched on/off by function 52", function
52 can also be used to address other service functions. These are restricted to service
engineers and are only visible when the service code is entered (code level 3).