
Technical data*
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
115
Table 7- 6
Technical data – measuring behavior
Measuring behavior
Output signal variation
< 1% of smallest possible measuring range
according to rating plate with an electronic delay
of 1 s
Measured value drift
< ± 2%/month of respective span
Repeatability
< ± 3 % of respective span
Linearity deviation
< ± 2% of respective span
Accuracy
5)
The accuracy is the geometric total of the output
signal variation, drift, repeatability and linearity
deviation with constant influencing variables.
Table 7- 7
Technical data – influencing variables
Influencing variables
3)
Ambient temperature
< 2%/10 K referred to the respective span
Sample gas pressure
With pressure compensation switched on:
< 0.2% of respective span /1% change in
pressure
Auxiliary gases
Catalytically inactive sensor:
approx. 1 vpm O
2
with an auxiliary gas
concentration of 10 vpm H
2
; CO and CH
4
exhibit
a smaller cross-interference; higher hydrocarbons
are negligible
Catalytically active sensor:
only gases with non-flammable auxiliary gas
components may be sampled!
Sample gas flow
< 2% of smallest possible span according to
rating plate with 10 ml/min change in flow within
the permissible flow range