Repair, maintenance and servicing
6
6.1
Important maintenance information
WARNING
Maintaining the device
Disconnect the gas and power supply before opening the device.
Only carry out setting work with appropriate tools in order to prevent short circuits on the
electronic boards.
Incorrect installation or adjustment could result in the release of gas, which in turn could
result in damage to the analyzer (e.g. danger of explosion) or to the health of persons (e.g.
poisoning).
Maintenance
The analyzer must be subject to annual maintenance in order to check the electrical safety
and correct working. It is up to the owner to decide on extension of the maintenance interval
in individual cases.
If standards or directives apply to the gases/components used, these must also be observed.
To allow maintenance work, the top cover can be removed and the front panel swung
forwards. If maintenance work is interrupted for more than two hours, the analyzer must be
closed again.