Operation
5.4 Operating functions
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
99
5.4.6.16
Error on/off (function 87)
87 Error on/off O
2
...continue
S1
Parameter memory
S2
S3
S4
External fault
Figure 5-67 Switching errors on and off
With this function, you switch the message for maintenance requests and faults (see Tables
"Causes of maintenance requests" and "Causes of fault messages") off individually so that
there is no entry in the logbook, no message in the status line of the measured value display
and no external signal sent.
Fault messages which do not apply to this analyzer are identified by the absence of text
following the fault number.
See also
Removing faults (Page 109)
5.4.6.17
PROFIBUS configuration (function 90)
Relay by PB :off:
9O Profibus Config. O
2
Address :126:
TAG : OXYMATSIXTYFOUR
Ident number : 1:
Software version: 2.0.O
Boot software : 0.2.0
Figure 5-68 Configuring PROFIBUS
This function can only be called if the device contains additional PROFIBUS electronics
(optional board).
With this function, you can set the PROFIBUS station address. The address can be set from
0 to 126.
Under TAG you can display the tag assigned to a device in the ELAN network (or the 16
characters thereof).