Byte no. Value
(hex)
Meaning
[90]
FF
Channel B, final value (-5,242,200 nm)
[91]
B0
[92]
02
[93]
A8
7.4.11
Error messages
Action in case of incorrect measurements
If a measurement cannot be run, an error is output. Possible causes of the error:
•
The measuring object is outside of the measuring range.
•
The light signal received by the device is not strong enough.
•
The laser is switched off.
You can configure the device behavior for the event that no measurement is possible.
The following options are available:
•
Error suppression time:
It is possible to set a time for which the last valid meas‐
ured value or the error value is displayed and held.
•
Hold last value:
If no measurement is possible, the last valid measured value is
displayed and held until a valid measured value is available again or for the dura‐
tion of the specified error suppression time.
•
Substitute value in the event of an error:
If no measurement is possible, the speci‐
fied replacement value is displayed and held until a valid measured value is avail‐
able again or for the duration of the specified error suppression time.
•
Substitute value in the event of an error
: A numerical value can be entered which
is output when no measurement is possible.
When an error occurs, the device sends an error message in the following format:
Table 43: Format of the error message
E0
02
Details of error
E0 02 Details of error
Table 44: Error messages
Name of error
Details of error
Details
Command error
0x00E1
The specified command does not
exist
Address error
0x00E2
The specified address exceeds the
permitted range
Overflow
0x00E9
Data overflow is returned
OPERATION
7
8021391//2017-10-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | OD5000
67