Aoralscan 2 User Manual
SHNING 3D TECH Co.,Ltd Copyright
2019 All Rights Reserved
- 14 -
Voltage
dips,
short
interruptions and
voltage
variations
on
power
supply
input lines
IEC 61000-4-11
0% UT (100% dip in
UT) for 0.5/1 cyclea
40% UT (60% dip in
UT) for 5 cycles
70% UT (30% dip in
UT)
for
25/30
cyclesa (for
0.5 sec)
0% UT (100% dip in
UT) for 250/
300 cyclesa
(for 0.5 sec)
0% UT (100% dip
in UT) for 0.5/1
cyclea
40% UT (60% dip
in UT) for 5 cycles
70% UT (30% dip
in UT) for 25/30
cyclesa (for
0.5 sec)
0% UT (100% dip
in UT) for 250/
300 cyclesa
(for 0.5 sec)
Mains power quality should
be
that
of
a
typical
commercial
or
hospital
environment. If the user of
the
Aoralscan 2
requires
continued operation during
power mains interruptions,
it is recommended that the
Aoralscan 2
be powered
from
an
uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location
in
a
typical
commercial
or
hospital
environment.
If image distortion occurs, it
may be necessary to
position the
Aoralscan 2
further from sources of
power frequency magnetic
fields or to install magnetic
shielding.
The
power
frequency magnetic field
should be measured in the
intended
installation
location to assure that it is
sufficiently low.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
a
For example, 10/12 means 10 cycles at 50 Hz or 12 cycles at 60 Hz.
Table 2-4 Guidance and manufacturer
’
s declaration
–
electromagnetic immunity
Summary of Contents for Aoralscan 2