12. Units
Read Chapter 7 Before Use of
These Pages
12.2 Percentage (%) Conversion
1
Set the % unit with menu item
if it is not set
up.
The % unit is set before shipment.
2
Press the
[UNIT]
key several times in the mass
display until the % unit is displayed.
Setting the 100% reference
1
Press the
[O/T]
key to tare the balance.
2
Load the reference sample that corresponds to
the 100% value. This value must be equivalent to
100 counts or more in the “g” unit.
3
When the stability mark illuminates, press the
[CAL]
key.
“SEt” is displayed briefly and the reference sam-
ple weight is displayed as 100%.
The weights of subsequent samples are dis-
played as a percentage of the reference sample
weight.
56
Note
The numbers of digits displayed in the % unit and rounding off of the mini-
mum digit vary depending on the mass value of the reference sample and
the balance model. It is not possible to obtain resolution greater than that in
the “g” unit.