14
4. Installation
For Basic Operation
3
This balance has three level screws (adjustable
feet) at the right front, left front and right rear cor-
ners.
Turning a level screw clock-wise stretches the leg
to raise the balance body there. Turning anti-
clockwise withdraws the leg and lowers the bal-
ance body.
The level indicator locates at left rear. The bubble
of it is off center when the balance is not placed
level.
(1) Adjustment is made with the two front level
screws only. Accordingly, first turn the right rear
level screw
1
anti-clockwise to withdraw its
leg completely.
(2) While adjusting level screws and observing
the bubble, gently press the left front corner of
the balance
2
so that both front level screw
feet
3
are touching the table surface.
(3) Bubble moves to the highest position. There-
fore, adjust level screws
3
so that the balance
main body is lowered in the direction of the
bubble.
Case 1: Right front of the balance is too high.
Turn right front level screw anti-clockwise so
that the bubble moves towards center.
Case 2: Front of the balance is too low. Turn
both front level screws clockwise so that the
bubble moves towards center.
(4) When the bubble has come to the center of
the red circle, turn the right rear level screw
clockwise until its foot softly touches the table
surface. Verify the balance sits stable with four
feet.
Level indicator
Level screw
(At the shortest
point when
starting
adjustment)
Level screw
Level
screw
Left front
3
3
2
1
[Large pan model]
Case 1
Case 2
Bubble
Bubble
Summary of Contents for UW2200HV
Page 2: ......
Page 8: ... VI ...
Page 9: ... VII ...
Page 10: ... VIII ...
Page 127: ...111 Appendices W Weighing 22 23 WindowsDirect 24 93 Z Zero range 62 Zero tracking 52 ...