42
10. Calibration
Read Chapter 7 Before Use of
These Pages
10.2.3 Span Calibration Using External Weights
Not applicable to a verified balance as a legal measuring instrument in the EU
The balance is adjusted using your external standard calibration weight(s).
1
Verify that the balance is in mass display and
that the pan is empty.
2
Press the
[CAL]
key once. “E-CAL” is dis-
played.
(If “E-CAL” is not displayed, return to mass dis-
play and select menu item
.)
3
Press the
[O/T]
key.
The value of the correct calibration weight to be
loaded is displayed and blinks.
4
Load the indicated calibration weight and press
the
[O/T]
key.
5
Shortly, zero display blinks. Unload the weight
from the pan and press the
[O/T]
key.
“SEt” is displayed briefly to indicate completion of
span calibration.
(Example)
3
Changing the Calibration Weight to be Used
Pressing the
[CAL]
key allows changes to the weight value. Modify the value
using the
[UNIT]
key and
[PRINT]
key, then press the
[O/T]
key. Refer to
for how to make numerical input. To interrupt modification, press the
[POWER]
key. Calibration range with external weights is designated to
each model. Refer to
for calibration range. Attempt of
inputting an invalid calibration weight value causes an error message.
Note
OIML Class E2 or F1 calibration weight is recommended for calibration,
depending on your accuracy demand.