70
14. Application Functions
Read Chapter 7 Before Use of
These Pages
Notes
• Press the
[POWER]
key in the peak detection standby state to initiate
the power standby state.
• Press the
[POWER]
key during detection of the peak to return to the
peak detection standby state.
• Peak Hold setting can be cleared without menu operation, by pressing
the
[POWER]
key for about three seconds.
• Polarity of the peak value displayed is “polarity of the displayed value of
the first change (by five times or more of Zero Range) from the display
within Zero Range.”
• Usually the peak value is easily measured by selecting menu item
(Pouring mode). Depending on the weighting conditions and the sample
type, this setting may not always be suitable.
• Pretaring Value (
) function cannot be used with Peak Hold function.
• See
for details of Zero Range.
• See
for details of stability detection.
• Not more than one of these seven functions can be set at a time. Setting
the second one with the menu will automatically clear the previously set
function; Auto Print(
), Interval
Timer(
lation mode(
).
• When minimum display has been switched to 10d (See
), the opti-
mized Zero Range for 1d display may cause failure in the operation.
Select optimal Zero Range value again when minimum display has been
changed (See
for details).
23