72
14. Application Functions
Read Chapter 7 Before Use of
These Pages
14.5 Add-on Mode
This function is convenient for weighing a large number of sample components as they are added on
the balance. Every time a component is added and stability is detected, the mass of it is displayed and
output through RS-232C or DATA I/O interface. The display will then be automatically zeroed for the
next component. When all the components have been weighed, the total mass is calculated by press-
ing
[POWER]
key.
Select menu item
.
1
Load the weighing vessel and press the
[O/T]
key in the Add-on standby state (The “Add-on”
and the Standby symbols are lit).
Zeroing occurs.
2
Press the
[PRINT]
key.
The Stand-by symbol disappears, and Add-on
measurement starts.
3
Load the first sample.
When stability is detected with a displayed value
five times the Zero Range or more, the displayed
value is automatically output and zeroing occurs.
Even if the displayed value is less than five times
the Zero Range, pressing
[PRINT]
key at stability
causes the same effect.
4
Repeat the above step
3
until all the compo-
nents have been weighed.
5
Press the
[POWER]
key.
The total sample mass on the pan is displayed
and the balance returns to the Add-on standby
state.
Press the
[PRINT]
key to output the total mass.
Start the next set of measurements from the
above step
1
.