Be sure that (the stability
mark) is displayed, and adjust the
zero point.
[0.000 ]
Before adjusting the zero point, be sure that
the pan supporter and an empty sample pan are
placed.
^
"Installing Parts", page 35
Do not expose the moisture analyzer to wind or vibration before adjustment is completed.
What is the zero point adjustment?
An operation that cancels the mass of tare such as sample pan to measure the mass of sample
correctly.
*
The heater cover during the zero point adjustment
Always be sure that the heater cover is closed when adjusting the zero point.
*
After adjusting the zero point.
Start measurement within thirty minutes after the zero-point adjustment. If you do not
start measurement within thirty minutes, a message (TIM.oUT) occurs in order not to do
unexpected operation.
If the message occurs, press
to return to the standby state (measurement display).
Start measurement promptly after adjusting the zero point.
5
Place a sample (object) on the
sample pan.
Be sure to place the sample as flat as possible
on the pan so that heat is applied evenly to the
sample during measurement.
Use a sample pan kept at room temperature.
^
"To Obtain the Best Results", page 47
*
A sample that is less than 0.0
grams cannot be measured.
A sample that is less than 0.02 grams
may be detected as a balance error.
Place a sample with at least 0.02 grams
on the sample pan.
! ! !0.0!0!0!
Check whether the stability mark is
displayed.
! ! !0.0!0!0!
Check whether the stability mark is
displayed.
! ! !5.0!1!2
! ! !5.0!1!2