MaINtENaNCE
When the heater has deteriorated or when the heating unit needs to be replaced due to the blow
of lamp, replace the heater with a new one.
Use the following heater replacement procedure.
WaRNINg
Instructions
Pull out the power cable from a receptacle. Otherwise, you may receive an electric shock.
WaRNINg
Instructions
Start the service only after the heater and the heater cover have cooled down. Doing so may
result in a burn injury.
*
Service life of heater
The halogen heater of the moisture analyzer deteriorates or its lamp may blow after approximately 5,000
hours although it depends on the application conditions of the moisture analyzer.
*
If you have replaced the heater…
The heater output may change. We recommend that you calibrate the temperature (
^
page 73
).
1
Remove the glass case.
^
"Removing the Glass Case", page 99
2
Unplug the connector of the used
heater cable.
3
Pull out the cables through the
cable supports.
Heater
Connector
Cable
supports
Heater
Connector
Cable
supports
Replacing
Heater