When These Messages Appear…
When any of the following messages appears on the panel, take actions given.
Message code
Possible cause
actions taken
See
ABoRT
(Operation aborted)
Span calibration or
measurement was aborted.
l
Press
to return to the standby
state.
l
Page 48
CoM.ERR
(Command error)
An unrecognizable command
code was received.
l
The error code appears and the
moisture analyzer enters the error
state.
Issue the correct command code.
l
Page 89
ERR.001
ERR.00
(Hardware error)
The temperature sensor has
failed.
The hardware information is
incorrect.
l
l
Unplug the power cord, and
then plug it in again. If the same
message appears again, call a
service representative.
l
ERR.005
(Memory error)
Memory has failed.
l
Unplug the power cord, and
then plug it in again. If the same
message appears again, call a
service representative.
l
ERR.100
The heater cover was left open
for more than one minute
during measurement.
l
Stop the measurement by pressing
.
l
Page 48
ERR.101
ERR.10
The heater temperature sensor
has failed.
l
Unplug the power cord, and
then plug it in again. If the same
message appears again, call a
service representative.
l
ERR.110
The heater lit is not closed.
l
Press
and close the heater
cover correctly.
l
Page 46
ERR.11
ERR.1
ERR.1
(Heater failure)
The heater cover or the heater
has failed.
l
Unplug the power cord, and
then plug it in again. If the same
message appears again, call a
service representative.
l
ERR.1
The ∆M value is not reached
four hours after measurement.
l
Recheck the ∆M value or the set
temperature.
l
Page 46
ERR.00
The power supply has failed.
l
Unplug the power cord, and
then plug it in again. If the same
message appears again, call a
service representative.
l
Page 38
ERR.01
The frequency judgment is
abnormal.
l
Unplug the power cord, and
then plug it in again. If the same
message appears again, call a
service representative.
l
Page 38
n