6
Measuring Moisture Content
s
2
MEaSuRE
6
Close the heater cover.
The measurement starts automatically.
*
If the measurement does not start
automatically…
Manual start is set for measurement.
Press
to start the measurement.
*
Switching between automatic and
manual start modes
As the default setting, measurement
start automatically after a sample is
placed and the heater cover is closed.
You can switch the mode to manual
start.
^
"Setting the Method to Start
Measurement", page 66
*
To check the ΔM…
When
is pressed in any
measurement mode, the display
switches and the ∆M value is shown.
Press this key again to return to the
previous display.
What is ΔM?
∆M is the rate of moisture change after 30 seconds.
^
"Setting Temperature and the Amount of Change in Moisture Content (∆M) as
Measurement Conditions (AUTO: Standard drying automatic ending mode)", page 53
*
Do not touch the inside of the heater cover immediately
The heat-dispersing component of the heater cover and the sample pan release heat during and
immediately after the measurement. Do not touch the moisture analyzer with your bare hands.
7
Check the measurement result.
Short beep sounds indicate that the measurement
is completed. The display panel shows the
measurement result and the measurement
completed indicator ( ). You can output
the measurement result to a printer or PC if
necessary. The factory default setting for the
output results is "FINAL".
^
"Setting a Measurement Data Printout
Timing", page 92
The heater fan stops automatically.
! !S!T!A!R!T
! ! ! !0.2!7
! !S!T!A!R!T
! ! ! !0.2!7
! ! ! !0.9!5
! ! ! !0.9!5