Preparing for Measurements
s
1
BEFORE MEaSuREMENt
The Span Calibration after Installing the Moisture Analyzer
After installing the moisture analyzer, we recommend you warm it up and perform span
calibration to stabilize the moisture analyzer. The span calibration is possible between 10 and 60
grams, however, a 50-gram calibration weight is recommended.
1
Turning the power on.
^
"Turning the Power On", page 38
2
Wait for at least one hour (to warm up the moisture analyzer).
To calibrate the sensitivity for precision measurement, the moisture analyzer must be in a stable
state wherever possible. To stabilize the moisture analyzer, we recommend you leave it to the
standby state for at least one hour. This is called "Warming-up".
3
Start the menu.
[
]
The menu appears.
4
Start the span calibration.
[CAL]
[BAL]
[WAIT]
[50.000 ]
If [TEMP] is displayed instead of [BAL], press
or
to display [BAL].
5
Open the heater cover.
n
!P!R! !G!R!M
!P!R! !G!R!M
! ! ! !C!A!L!
! ! ! !B!A!L
! !5!0.0!0!0!
/ / / / /
/ /
/
//
//
//
//
//
//
//
/
/ /
/ / / /
/ /
! ! ! !C!A!L!
! ! ! !B!A!L
! !5!0.0!0!0!
/ / / / /
/ /
/
//
//
//
//
//
//
//
/
/ /
/ / / /
/ /