Place a 50-gram calibration weight
on the sample pan.
*
When you use a weight other than
50 grams...
If you use a weight other than the
50-gram one, reset the corresponding
mass by pressing
. Then,
press
and place the weight.
When the weight is placed, the measurement of
50 grams starts.
*
To Calibrate with the heater cover
closed…
Ensure that the glass protect plate
does not come in contact with the
calibration weight when the heater cover
is closed. If it does, open the heater
cover and perform calibration within an
environment without any disturbance
(vibration, wind).
7
Lower the calibration weight when
[0.000] starts to flash.
The measurement of 0 grams starts.
8
Wait until [END] is displayed.
The span calibration ends when [END] is
displayed.
The display returns to BAL after a short period
of time. Press
to enter the standby state.
You can also calibrate the temperature. For
further information, see
"Calibrating the
Temperature (Option)" (
^
page 73)
.
When the mass indication flashes,
place the weight.
! !5!0.0!0!0!
/ / / / /
/
/
/
//
//
//
//
//
//
/
/
/ /
/ / / / /
When the mass indication flashes,
place the weight.
! !5!0.0!0!0!
/ / / / /
/
/
/
//
//
//
//
//
//
/
/
/ /
/ / / / /
When the mass indication flashes,
remove the weight.
! ! !0.0!0!0
/ / / / / /
/
/
/
//
//
//
//
//
//
/
/
/ /
/ / /
When the mass indication flashes,
remove the weight.
! ! !0.0!0!0
/ / / / / /
/
/
/
//
//
//
//
//
//
/
/
/ /
/ / /