Digital Weighing Indicator
SI 4100
G
TGX]GTG
Ex) Load Cell CAPA : 20kg, division 0.001
Use test weight(at least 2kg) which is at least over 10% of
max CAPA(20kg)
input test weight 2.000
ll 2.000
G
13. After displaying “UP” ,please load “Test Weight” .
Ex) Load Cell CAPA : 20kg, division 0.001
Use test weight(at least 2kg) which is 10% of max CAPA(20kg)
Up
14. And press
key.(Do not unload test weight)
Up
15. Indicator will calculate span value during 5sec, automatically
Cal-2 0
16. After calculation, span value displays on the display.
Please unload the test weight.
0.629238
G
Caution
ଖ
:
The
“Test Weight’s value”
must be at least
10%
max. capacity of weighing scale.
“at least 10%” means to guarantee precise weighing process you have to make standard with at least
10% of weight of Max. capacity.
We programmed the calibration will not be done, when you load less than 10% of max. capacity.
17. Press
key to save all calibration process.
After then it resets automatically.
(Now, fasten the Calibration Bolt. )