Digital Weighing Indicator
SI 4100
G
TGYYGTG
G
Zero key Operation Range selection
F08
0
Activated within 2% of Max. Capacity
1
Activated within 5% of Max. Capacity
̻
̻G
2
Activated within 10% of Max. Capacity
3
Activated within 20% of Max. Capacity
4
Activated within 50% of Max. Capacity
5
Activated within 100% of Max. Capacity
6
Whenever Press “Zero” key (No Limit)
Tare key Operation Range selection
F09
0
Activated within 10% of Max. Capacity
1
Activated within 20% of Max. Capacity
̻G
2
Activated within 50% of Max. Capacity
G
3
Activated within 100% of Max. Capacity
“Hold” Mode selection
F10
̻G
0
Peak Hold : Measure Max. weight value and hold on display.
G
1
Sample Hold : Hold current weight until “Hold Reset”.
External Input Selection
F11
Set Value
Input 1
Input 2
Input 3
Input 4
̻G
0
ZERO
TARE /TARE reset
Hold / Hold reset
G
1 RUN/STOP TARE
/
TARE
reset
ZERO PRINT
G
2 RUN
STOP
SUB-Print GRAND
3
STOP
TARE
TARE
RESET
4
RUN
STOP
TARE
TARE
RESET
5
RUN
STOP
ZERO
6
Zero
TARE
/
TARE
reset
HOLD HOLD
RESET
“STEADY” condition check time setting
F12
03
01
∫
20
During the set time period, estimate weighing part’s “STEADY” condition and display.
If you set small value, indicator will take “STEADY” fast, if you set large value,
indicator will take “STEADY” slow.