Digital Weighing Indicator
SI 4100
G
TGYXGTG
5-5. F-Function
■ General Function Setting
(“●” Factory default set value)
Weighing Data Save Method Selection
(Apply on Accumulated weighing count/weight)
F01
̻
̻G
0
Manual Save Mode (Save when “Print” key input)
G
1
Automatic Save Mode(Save when Finish relay Output)
2
Combined Save Mode (Save when “Print” key input or weighing is Finished)
3
Non-Save Mode(Any kinds of weighing data will not be saved).
Weight-Back up selection
F02
̻G
0 Normal
Mode
G
1
Weight Back up Mode
Motion Band Range setting
F03
06
0
∫
50
This is set “Steady” acceptable range of weighing part.
If there is vibration on weighing part, you can set this function and reduce the vibration
effect on weighing process.
0 : Weak vibration ~ 50 : Strong Vibration
Zero Tracking Compensation Range setting
F04
2
0
∫
9
Due to external causes(Temperature, wind, and dust), there are small weight difference,
indicator will ignore the weight difference and display Zero.
For this compensation function, indicator will estimate the weight difference is over the
set range during fixed time period.
If there is large weight difference over set range within fixed time period, the “Zero” is
breaking and will find new zero point.
Auto Zero Range setting
F05
00
00
∫
99
Within the “Auto Zero” range, weighing part is steady, indicator will display current
weight as “Zero”
If the weighing part is not “Steady”, indicator will display current weight.
(Auto Zero Range : ± Set value + weight unit)
Digital Filter setting
F06
13 AB
A : Frequency Filter setting value (0~1)
(0 : about 200Hz/sec, 1 : about 500Hz/sec)
B : Buffer Filter setting value (1~9)
If “B” set value is fixed, “A” set value is
large, the indicator will response more
sensitive.
Zero /Tare key Operation mode selection
F07
̻G
0
Activate when “Steady” condition, only
G
1 Always
activated