5. SET-UP
5-1. Calibration
Adjust weight balance between “Real weight” on the load cell(Weight Part) and “Displayed weight of
Indicator”. When you replace LOAD CELL or Indicator, you have to do Calibration process once again
5-2. Test Weight Calibration Mode (Using Test weight)
ଖ
remarks : In case that
ˈ
P-W
ˉ
is displayed, you have to check the pass word.
Prepare At least 10% of Max. capacity of your weighing scale, and remove
“CAL-BOLT”
on the Front
panel and press
“CAL - LOCK S/W”
inside.
ଖ
remarks : In case that
ˈ
P-W
ˉ
is displayed, you have input the pass word to start
calibration mode.
G
1. At normal mode, remove “CAL-BOLT” on the Front panel
1123
2. And press “CAL - LOCK S/W” inside.
Check the “SET-CAL. Massage on display.
5et-Cal
ଖ
For the save the each step, press
key, for the cancel or move back, press
key.
3. If you press
key, Calibration Mode starts.
After displaying “CAPA”,
C999999
4. Please input Max capacity of your weighing scale,
And press
key.
Ex) Load cell CAPA : 20kg, division : 0.001
Input 20000
C 20000
5. define the optimal position or Decimal point
Whenever pressing
key, Decimal point will be changed.
Ex) Load Cell CAPA : 20kg, division : 0.001kg
input 20.000
p 20.000